Comparative Evaluation of Artifacts Originated by Four Different Post Materials Using Different CBCT Settings

MakaleAçık ErişimScopus SJRQ2
  • SDG 3 · Sağlık ve Kaliteli Yaşam

Özet

The aim of this study was to evaluate whether cone beam computed tomography (CBCT) images in the presence of four different post materials, obtained from different kVps with varying resolutions and varying metal artifact reduction (MAR) algorithms, differed in artifact estimation, and to compare tooth regions in terms of artifact value. Materials and Methods: Forty premolar teeth were used in this study. Root canals were treated, and teeth were randomly distributed into four subgroups (n = 10) for the preparation of post materials: titanium, gold (Nordin), quartz fiber (Bisco DT Light), and glass fiber (Rely X). The CBCT images were taken with two different kVps, three different metal artifact reduction (MAR) algorithm options, and two different resolutions. For each protocol, the effective dose was calculated according to the dose area production (DAP) value. The standard analysis of variance technique and the Tukey multiple comparison adjustment method were used to assess interactions among material types, kVp, MAR, and voxel settings. Results: More artifacts were found in the middle third than in the cervical third (p < 0.05). The mean value of artifacts was highest for gold (Nordin), 90 kVp, no MAR, and 100 voxel size. Glass or quartz fiber posts at low resolution, with high MAR and 96 kVp, originated fewer artifacts. Moreover, the use of 90 and 96 kVp with 200 voxel size and high MAR provided the least amount of radiation. Conclusion: The best setting for radiographic follow-up of post materials on the Planmeca ProMax is 96 kVp with low resolution and high MAR; this setting produced one of the lowest effective doses. Clinical Significance: This study estimated the best scanning protocol by lowering the effective dose to a minimum level according to the “as low as reasonably achievable” principle, as well as assessing the tooth region and the post material generating the fewest artifacts, in order to prevent image interpretation challenges such as false-positive and false-negative results stemming from the deterioration of the visibility of the root canal due to perforation, fractures, and voids in the root canal region.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/kostu/10431Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.3390/tomography8060245Yayıncı resmi DOI adresi
İndeks kayıtları
Kaydın indeks veri tabanlarındaki sayfası (aktarımda alınan kayıt numarasıyla). Erişim, indeksin kendi koşullarına bağlıdır.
Atıflar
Scopus
7
WoS
4
OpenAlex
5
Her kaynağın kendi saydığı toplam atıf; aktarım ve zenginleştirme sırasında alındı.
Tam Kayıt
AlanDeğerDil
dc.contributor.authorHelvacioglu-Yigit, Dilek
dc.contributor.authorSeki, Umut
dc.contributor.authorKursun-Cakmak, Sebnem
dc.contributor.authorDemirturk Kocasarac, Husniye
dc.contributor.authorSingh, Maharaj
dc.date.accessioned2026-10-27T23:47:32.039204605Z
dc.date.available2026-10-27T23:47:32.039204605Z
dc.date.issued2022
dc.description.abstractThe aim of this study was to evaluate whether cone beam computed tomography (CBCT) images in the presence of four different post materials, obtained from different kVps with varying resolutions and varying metal artifact reduction (MAR) algorithms, differed in artifact estimation, and to compare tooth regions in terms of artifact value. Materials and Methods: Forty premolar teeth were used in this study. Root canals were treated, and teeth were randomly distributed into four subgroups (n = 10) for the preparation of post materials: titanium, gold (Nordin), quartz fiber (Bisco DT Light), and glass fiber (Rely X). The CBCT images were taken with two different kVps, three different metal artifact reduction (MAR) algorithm options, and two different resolutions. For each protocol, the effective dose was calculated according to the dose area production (DAP) value. The standard analysis of variance technique and the Tukey multiple comparison adjustment method were used to assess interactions among material types, kVp, MAR, and voxel settings. Results: More artifacts were found in the middle third than in the cervical third (p < 0.05). The mean value of artifacts was highest for gold (Nordin), 90 kVp, no MAR, and 100 voxel size. Glass or quartz fiber posts at low resolution, with high MAR and 96 kVp, originated fewer artifacts. Moreover, the use of 90 and 96 kVp with 200 voxel size and high MAR provided the least amount of radiation. Conclusion: The best setting for radiographic follow-up of post materials on the Planmeca ProMax is 96 kVp with low resolution and high MAR; this setting produced one of the lowest effective doses. Clinical Significance: This study estimated the best scanning protocol by lowering the effective dose to a minimum level according to the “as low as reasonably achievable” principle, as well as assessing the tooth region and the post material generating the fewest artifacts, in order to prevent image interpretation challenges such as false-positive and false-negative results stemming from the deterioration of the visibility of the root canal due to perforation, fractures, and voids in the root canal region.
dc.description.sponsorshipKocaeli Üniversitesi, (HD2017/70)
dc.description.sponsorshipKocaeli Üniversitesi
dc.identifier.doi10.3390/tomography8060245
dc.identifier.eissn2379-139X
dc.identifier.endpage2928
dc.identifier.issn2379-1381
dc.identifier.issue6
dc.identifier.pmcidPMC9781935
dc.identifier.pmid36548537
dc.identifier.scopus2-s2.0-85144487633
dc.identifier.startpage2919
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/kostu/10431
dc.identifier.volume8
dc.identifier.wosWOS:000902978800001
dc.language.isoen
dc.publisherMDPI
dc.relation.ispartofTomography
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectartifacts
dc.subjectbeam hardening
dc.subjectcone-beam computed tomography
dc.subjectdental post
dc.subject.classificationRadiology, Nuclear Medicine & Medical Imaging
dc.subject.otherArtifacts
dc.subject.otherCone-Beam Computed Tomography
dc.subject.otherGold
dc.subject.otherQuartz
dc.subject.otherSpiral Cone-Beam Computed Tomography
dc.subject.othersilicon dioxide
dc.subject.otherartifact
dc.subject.othercone beam computed tomography
dc.subject.otherprocedures
dc.titleComparative Evaluation of Artifacts Originated by Four Different Post Materials Using Different CBCT Settings
dc.typeArticle
rdl.citation.openalex5
rdl.citation.scopus7
rdl.citation.wos4
rdl.departmentCollege of Dental Medicine, QU Health, Qatar University, Doha 2713, Qatar
rdl.departmentDepartment of Dentomaxillofacial Radiology, Faculty of Dentistry, Kocaeli University, Kocaeli 41190, Turkey
rdl.departmentDepartment of Dentomaxillofacial Radiology, Faculty of Dentistry, Ankara Yıldırım Beyazıt University, Ankara 1487, Turkey
rdl.departmentBeamReaders Inc., Kennewick, WA 99336, USA
rdl.departmentDepartment of Dentomaxillofacial Radiology, Kocaeli Health and Technology University, Kocaeli 41000, Turkey
rdl.departmentCollege of Nursing, Marquette University, Milwaukee, WI 53233, USA
rdl.oa.statusgold
rdl.oa.statusgreen
rdl.oa.urlhttps://www.mdpi.com/2379-139X/8/6/245/pdf?version=1670920449
rdl.scopus.quartileQ2
rdl.sdgSDG03
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.source.indexPUBMED
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)